Trade name: Technical name: Registration number. . Ammonia Reagents (AMM); Ammonia 10345161867. . Transaminase Glutamic P-Transaminase 10345160498. Rubic (TGP OR ALT); Pyruvic Glutamic. (Tgp or Alt). . . ADVIA CHEMISTRY Alanine Transaminase 10345161103. Aminotransferase Pyruvic Glutamic Concentrate. ted Reagents (ALT_C); (TGP or ALT). . . Dimension Amonia Flex; Ammonia 10345161934. . Alanine Amino Transaminase Reagent 10345161852. transferase ALTI (TGP) Tell me Pyruvic Glutamic. nsion; (TGP or ALT). . . . Aspartate Amino Transaminase Reagent 10345161537. transferase (AST) - Dimension; Glutamic. (TGP or ALT). . . Glucose Reagent Glucose Glucose 10345161260. (Gluc) Dimension ;. . . Dimension MBI Flex; Isoenzyme Activity 10345161715. Creatinine kinase. . Thyroxine Reagent (T4) Thyroxine (T4) 10345161355. Size. . Risk class: II. . Affected Model:. . Serial numbers affected: N / A. 의 안전성 경고

Agência Nacional de Vigilância Sanitária (ANVISA)에 따르면, 해당 안전성 경고 는 Brazil 에서 Siemens Healthcare Diagnósticos S.A 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

안전성 경고는 의료기기 제품과 관련된 중요한 정보와 권고사항을 담고 있습니다. 물론 안전성 경고가 배포되었다고 해서 해당 제품이 무조건 안전하지 않은 제품이라는 것은 아닙니다. 보건의료업계 종사자와 의료기기 사용자들에게 배포되는 안전성 경고에는 회수(recall)도 포함될 수 있습니다. 제조사가 안전성 경고를 작성하기도 하지만, 보건당국에서 작성하는 경우도 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert
  • 사례 ID
    2343
  • 날짜
    2017-07-21
  • 사례 국가
  • 사례 출처
    ANVISA
  • 사례 출처 URL
  • 비고 / 경고
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • 데이터 추가 비고
    It is recommended that venipuncture should occur prior to the administration of sulfasalazine or sulfapyridine. If you wish to notify technical complaints and adverse events use the following channels: Notivisa: Notices of adverse events (AE) and technical complaints (QT) for products subject to Sanitary Surveillance should be made through the NOTIVISA System. To access the System, you must register and select the Health Professional option, if you are a liberal professional or the Institution / Entity option, if you are a professional of an institution / entity. Technovigilance System: Patient or citizen can notify through the Technovigilance System / SISTEC access through the link - Date of identification of the problem by the company: 06/29/2017 - Date of notification notice to Anvisa: 07/21/2017 The company that owns the affected product is responsible for contacting its customers in a timely manner to ensure the effectiveness of the Field Action in progress. It stands out the joint responsibility of the distribution chain and use of health products in maintaining its quality, safety and efficacy, as well as the effectiveness of the Field Action, expressed by RDC 23/2012: "(...) Art 2 ° A holder of registration of health product is the holder of the registration / registration of health product with Anvisa. Single paragraph. The registration holder, as well as other agents involved from the production to the use of the product, or discarding it when applicable, are jointly and severally responsible for maintaining the quality, safety and efficacy of the health products to the final consumer. Art. 12 Distributors of health products shall forward to the registration holder, in a timely manner, the distribution map and other information requested for the notification and execution of field actions. (...) "
  • 원인
    Siemens healthcare diagnostics has identified possible sulfasalazine and sulfapyridine interference in assays using nadh and / or nadph to generate oxidation reduction reactions producing colorimetric signals.
  • 조치
    Field Action Code CHC 17-06 triggered under the responsibility of the company Siemens Healthcare Diagnósticos SA Issue letter to the client.

Manufacturer