원인
A potential problem which could result in an early failure of the scissor arm to bracket connection. the failure can be identified by observing the physical condition of the scissor arm bracket. the plastic covers around the bracket typically show signs of fatigue before the bracket actually breaks. this can be an early indicator of a potential problem.