원인
The master pumps used to calibrate manufactured and serviced pumps have a calibration shift of up to 6.8% (+3%). when an affected master pump is used to calibrate production and serviced pumps this pump is used to calibrate production and serviced pumps this calibration shift could lead to over delivery of fluid beyond the label claim of +5%.