Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Defective wheels on wheeled walker. the cause was that the wheel hub thickness went from 5mm to 4mm. this change causes some hubs to fracture due to side impact or heavy weight leaning to one side of walker.
Model Catalog: (Lot serial: 07722-8); Model Catalog: (Lot serial: 07722-5); Model Catalog: (Lot serial: 07725); Model Catalog: (Lot serial: 07722G); Model Catalog: (Lot serial: 30757WB); Model Catalog: (Lot serial: 30751W); Model Catalog: (Lot serial: 30757W); Model Catalog: (Lot serial: 30758W)