원인
Tantalum capacitor c5 was installed backwards during assembly which may result in premature failure. failure may result in immediate reset of the monitor or shutdown of the device without an alarm.
Model Catalog: 6004013 (Lot serial: SN AJ07080006); Model Catalog: 6004013 (Lot serial: SN AJ07080007); Model Catalog: 6004013 (Lot serial: SN AJ08010004)