Beckman coulter determined that the quality control (qc) value card lists incorrect standard deviation (sd) values for the access amh qc kit lot 689202.The qc value cards for this lot list the following standard deviation values: qc1 as 0.02 qc2 as 0.11 and qc3 as 0.65. the correct standard deviation values are as follows: qc1 is 0.15 qc2 is 0.76 and qc3 is 2.30.