Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Beckman coulter has determined that the access 2 console pc (b23083) may experience an "mfc exception" error during normal operation. if this error occurs the system will exit the access 2 user interface (ui) screen and display a window that notifies the user of a "0==itubescancount" event. if the ui is re-launched the system will prevent access to the sample manager screen to request samples for processing and the same "mfc exception" error will occur.