원인
Beckman coulter has determined through internal testing that access ostase calibrator lots 528936 528938 and access ostase qc lots 528937 528939 do not meet their labelled expiration dates when stored after initial use (open vial stability). beckman coulter has determined that the above lots are stable at 2 to 10 degrees celcius for 58 days after initial use (i.E. after the vials are first opened). when stored unopened calibrators and quality controls are stable until the expiration date on the vials.