원인
Impaction forces may fracture the welds holding the striker plate to the rasp handle. this may lead to fatigue stress fractures and disassociation of the strike plate from the rasp handle. the surgeon may proceed with the surgery by removing and replacing the brocken rasp handle. an extension of surgery time of around 30 minutes is required to fix the device.