Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Potential corruption of the meter's flash memory caused by a flipped bit. the manifested malfunctions of the flipped bit were a complete shutdown of the meter or a too-low and non-plausible result.