ACTIVA RC NEUROSTIMULATOR - DBS EXTENSION KIT FOR DEEP BRAIN STIMULATION 의 리콜

Health Canada (via FOI)에 따르면, 해당 리콜 는 Canada 에서 MEDTRONIC OF CANADA LTD. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    111327
  • 사례 위험등급
    II
  • 사례 시작날짜
    2016-05-17
  • 사례 국가
  • 사례 출처
    HC
  • 비고 / 경고
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • 데이터 추가 비고
  • 원인
    Notification regarding information that will be added in the future to medtronic's deep brain stimulation (dbs) labeling. the labeling updates further clarify potential risks which have been reported with dbs therapy. medtronic is sharing this information to help with management of current and future patients treated with dbs therapy. medical safety assessment evaluated reports of a loss of swimming ability or drowning following dbs implant. the dbs manuals reviewed as part of the msa identified language describing the potential for loss of coordination however this language does not describe the potential for changes in swimming ability due to dbs therapy and the associated risk of drowning in the dbs patient population. medical safety assessment (msa) evaluated reports of serious injury associated with tunneling for dbs therapy. the dbs manuals reviewed as part of the msa do not describe some of the risks associated with tunneling identified in the msa which include spinal accessory nerve injury vascular injury tunneling too deeply too superficially or tunneling through unintended anatomy (such as in between the ribs or entering the thoracic cavity). note: the tunneling tool model associated with these injuries was not reported for some of these events.

Device

  • 모델명 / 제조번호(시리얼번호)
    Model Catalog: 37086 (Lot serial: All Lots); Model Catalog: 7428 (Lot serial: All Lots); Model Catalog: 37603 (Lot serial: All Lots); Model Catalog: 37602 (Lot serial: All Lots); Model Catalog: 37612 (Lot serial: All Lots); Model Catalog: 37601 (Lot serial: All Lots); Model Catalog: 7426 (Lot serial: All Lots); Model Catalog: 7483 (Lot serial: All Lots)
  • 제품 설명
    ACTIVA RC NEUROSTIMULATOR - DBS EXTENSION KIT FOR DEEP BRAIN STIMULATION;KINETRA NEUROSTIMULATOR PULSE GENERATOR & ACCESSORY KIT;ACTIVA SC - NEUROSTIMULATOR;ACTIVA PC INS - DBS EXTENSION KIT FOR DEEP BRAIN STIMULATION;ACTIVA RC MULTI-PROGRAM RECHARGEABLE
  • Manufacturer

Manufacturer

  • 제조사 주소
    BRAMPTON
  • 제조사 모회사 (2017)
  • 제조사 의견
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.