원인
To notify users of the acuson s family ultrasound systems of a recent finding regarding needle guides. two transducers 6c1hd and 18l6 hd demonstrate misalignment while utilizing the on-screen guidelines. the issue occurs when using the civco biopsy attachments in combination with the acuson s family on-screen guidelines. the biopsy needle may traverse outside the on-screen guidelines. this occurs on the 6c1 hd and 18l6 hd transducers.