원인
A single digital encoder cable assembly during inspection had a loose wire further investigation revealed a deviation in the height of a wire crimp could lead to a loose wire.
Model Catalog: 816570 (Lot serial: #5950 TO #5953); Model Catalog: 816571 (Lot serial: #5950 TO #5953); Model Catalog: 816570 (Lot serial: #6621 TO #6640); Model Catalog: 816571 (Lot serial: #6621 TO #6640)
제품 설명
ADVANCED PERFUSION SYSTEM 1 4"SMALL ROLLER PUMP AND 6"LARGE ROLLER PUMP ENCODER CABLE