원인
Siemens healthcare diagnostics is conducting a field corrective action for advia 120 hematology systems. the advia 120 optics cover or hood is located on the very top of the advia 120 analyzer. when the optics cover is raised it is supported by two gas struts which are attached to the cover on the left and right sides. over time it is possible that these struts may lose their effectiveness and fail to support the optics cover in the open position. this could cause the cover to fall during maintenance procedures.