원인
Siemens healthcare diagnostics has become aware of n-acetylcysteine (nac) and metamizole (dipyrone) interference with trinder and trinder-like reaction assays. the trinder reaction is a reaction where hydrogen peroxide is formed and subsequently reacts with a phenol derivative and aminoantipyrine in the presence of peroxidase to form a coloured quinone product. siemens has confirmed that falsely depressed results may occur on samples drawn from patients receiving n-acetylcysteine (nac) or metamizole. nac is the accepted antidote for acetaminophen toxicity and is justified in patients at significant risk for hepatotoxicity. metamizole is an anti-inflammatory anti-pyretic drug banned in most countries because of the potential for nephrotoxicity. the instructions for use (ifu) for the assays will be updated to indicate that: venipuncture should occur prior to n-acetylcysteine (nac) or metamizole administration due to the potential for falsely depressed results.
Model Catalog: 10361941 (Lot serial: all lots); Model Catalog: 04862501 (Lot serial: all lots); Model Catalog: 04992596 (Lot serial: all lots); Model Catalog: 10376501 (Lot serial: all lots); Model Catalog: 07497014 (Lot serial: all lots); Model Catalog: 03051305 (Lot serial: all lots); Model Catalog: 10335892 (Lot serial: all lots); Model Catalog: 07109944 (Lot serial: all lots); Model Catalog: B01-4818-01 (Lot serial: all lots); Model Catalog: 09796248 (Lot serial: all lots); Model Catalog: B01-4760-01 (Lot serial: all lots); Model Catalog: 07511947 (Lot serial: all lots); Model Catalog: 06521559 (Lot serial: all lots); Model Catalog: 10335872 (Lot serial: all lots); Model Catalog: 10697575 (Lot serial: all lots); Model Catalog: 10492319 (Lot serial: all lots); Model Catalog: 04993681 (Lot serial: all lots); Model Catalog: 04993452 (Lot serial: all lots); Model Catalog: 04993223 (Lot serial: all lots)
제품 설명
ADVIA 1200 CHEMISTRY SYSTEM - FRUCTOSAMINE (FRUC) REAGENT;ADVIA CHEMISTRY SYSTEM - ENZYMATIC CREATININE_2 (ECRE_2) ASSAY;ADVIA 1200 CHEMISTRY SYSTEM - CHOLESTEROL_2 (CHOL_2) ASSAY;ADVIA CHEMISTRY SYSTEM - URIC ACID (UA) ASSAY;ADVIA 1200 CHEMISTRY SYSTEM -