Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Siemens healthcare diagnostics is conducting a recall for advia 2120/2120i hematology systems that are connected to avdia autoslide systems (smn 10286141 or smn 10286142). siemens has determined that if an autosampler rack jam error occurs during operation of an advia 2120/2120i connected to an advia autoslide it is possible that the next slide processed by the autoslide could be labeled with the wrong sample identification information.