데이터 추가 비고
원인
The high calibrator for calibrator a lot ca77 is under-recovering to the assigned target for the advia centaur total t4 assay. internal investigations showed that the average bias of the high calibrator-to-target is -20% with an observed range of bias from -21% to -19%. the under-recovery of the high calibrator is leading to an elevated response in the high end of the total t4 assay which may be reflected in level 3 quality controls (15.8ug/dl for biorad lyphochek control level (40243)) being out of range high. internal investigations showed that the average bias of the level 3 control-to-target is 16% which had an observed range of bias from 10% to 20%.