ADVIA CENTAUR CP SYSTEM - CALIBRATOR 의 리콜

Health Canada (via FOI)에 따르면, 해당 리콜 는 Canada 에서 SIEMENS HEALTHCARE LIMITED 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    127762
  • 사례 위험등급
    III
  • 사례 시작날짜
    2013-07-03
  • 사례 국가
  • 사례 출처
    HC
  • 비고 / 경고
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • 데이터 추가 비고
  • 원인
    The high calibrator for calibrator a lot ca77 is under-recovering to the assigned target for the advia centaur total t4 assay. internal investigations showed that the average bias of the high calibrator-to-target is -20% with an observed range of bias from -21% to -19%. the under-recovery of the high calibrator is leading to an elevated response in the high end of the total t4 assay which may be reflected in level 3 quality controls (15.8ug/dl for biorad lyphochek control level (40243)) being out of range high. internal investigations showed that the average bias of the level 3 control-to-target is 16% which had an observed range of bias from 10% to 20%.

Device

  • 모델명 / 제조번호(시리얼번호)
    Model Catalog: 04800840 (Lot serial: 90705A77); Model Catalog: 04800735 (Lot serial: 90705A77); Model Catalog: 04800735 (Lot serial: 90669A77); Model Catalog: 04800840 (Lot serial: 90669A77); Model Catalog: 04800840 (Lot serial: 89983A77); Model Catalog: 04800735 (Lot serial: 89983A77)
  • 제품 설명
    ADVIA CENTAUR CP SYSTEM-CALIBRATOR