Siemens healthcare diagnostics has confirmed that samples containing fluorescein may show interference with the advia centaur systems tsh3 ultra vitamin d and the advia centau brahms procalcitonin assays. samples containing fluorescein can produce falsely depressed values when tested with advia centaur tsh3 ultra assay. samples containing fluorescein can produce falsely depressed values when tested with the advia centaur brahms procalcitonin assay. samples containing fluorescein can produce falsely elevated values when tested with the advia centaur vitamin d assay.