원인
Siemens healthcare diagnostics received customer feedback regarding the clinical utility of advia centaur brahms procalcitonin (pct) quality control level 1. quality control level 1 currently has a range of +/- 50% of the target. siemens internal investigation confirmed that this range was incorrectly set. siemens has revised the range of advia centaur brahms procalcitonin (pct) quality control level 1 to +/- 30% of the target. a review of test data confirmed the assay is capable of achieving a narrower range of +/- 30% of the target.