Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
An urgent field safety notice (ufsn) has been issued by siemens informing the customers of the biotin interference in the advia centaur tni-ultra assay. in date lots of advia centaur tni-ultra exhibit a greater than 10% change in results in samples with biotin levels up to 10 ng/ml (41 nmol/l). the instructions for use states that specimens that have up to 10 ng/ml (41 nmol/l) of biotin demonstrate =10 %change in results.