원인
Manufacturer has identified an error on the advia centaur confirmatory card that is included in advia centaur hbsagii reagent lot 109056 (kit lots 55825056 56167056 and 56244056). the error causes an interpretation of 'invalid' and a flag of 'conf n/a' when the hbsagii positive control is tested with the hbsagii confirmatory (conf) assay.