원인
Due to a design defect the 3-way valves may fail allowing cleaning solution which contains bleach to leach into the analytical lines and cause erroneous results.
Model Catalog: 04911855 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: 09066916 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: 078-A010-07 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: 078-A001-14 (Lot serial: ALL SERIAL NUMBERS)