원인
Siemens healthcare diagnostic has confirmed through internal investigation of lot 084 that the solid phase in some of the lot 084 tni-ultra readypacks is darker in appearance.
Model Catalog: 02789602 (Lot serial: KIT LOTS ENDING IN 084); Model Catalog: 02789602 (Lot serial: KIT LOTS ENDING IN 078); Model Catalog: 02789602 (Lot serial: KIT LOTS ENDING IN 083); Model Catalog: 02790309 (Lot serial: KIT LOTS ENDING IN 082); Model Catalog: 02789602 (Lot serial: KIT LOTS ENDING IN 082); Model Catalog: 02790309 (Lot serial: KIT LOTS ENDING IN 079); Model Catalog: 02789602 (Lot serial: KIT LOTS ENDING IN 079); Model Catalog: 02790309 (Lot serial: KIT LOTS ENDING IN 078); Model Catalog: 02790309 (Lot serial: KIT LOTS ENDING IN 084); Model Catalog: 02790309 (Lot serial: KIT LOTS ENDING IN 083)
제품 설명
ADVIA CENTAUR XP SYSTEM - TROPONIN I ULTRA (TNI-ULTRA) ASSAY