원인
Quality control (qc) is shifting higher impacting direct bilirubin (dbil_2) with siemens chemistry calibrator lots 090445b and 090445c from previous lots 090445 and 090445a. observed qc shifts are ~7-9% higher with calibrator lots 090445b and 090445c. an urgent field safety notice has been issued by the manufacturer to provide affected customers with reassigned system specific values for the dbil_2 ssvs for chemistry calibrator lot numbers 090445b and 090445c. manufacturer has determined that the dbil_2 ssvs currently assigned to these lots may cause an average shift of +7-9% in quality control and sample recovery at a level of 3 mg/dl. the urgent field safety notice provides reassigned values for the dbil_2 ssvs for chemistry calibrator lot numbers 090445b and 090445c.