Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Recent routine internal fire testing to test flammability after multiple washings of the device (n=5) of the whole cushion (cover and foam) in water after 40 degrees celsius revealed a potential risk of flammability from the product (the "testing failure"). this testing failure contradicts the information contained in the company's instructions for use wherein it states: "hand wash all foam breathair components in warm water at 40 degrees celsius using a standard mild detergent. dno not use fabric softner. rinswe well.".