데이터 추가 비고
원인
During beam delivery tolerance values are used to do checks of any unexpected movements. the problem is that the static tolerances from the calibration files (loaded database) are used for beam delivery instead of the machine calculated values. a fault will only occur if the loaded database is incorrect.