원인
The y adapter within the circuit may spontaneously crack causing a leak in the closed ventilation system the leak may cause a loss of tidal volume being delivered to the patient and the ventilator alarm may sound. the leak could lead to a decrease in the tidal volume delivered to the ventilated patient.
Model Catalog: 5925-504 (Lot serial: ALL LESS THAN LOT #000415257); Model Catalog: 5925-504 (Lot serial: ALL STARTING WITH THE LETTER Y); Model Catalog: 5925-504 (Lot serial: OR); Model Catalog: 1209-503 (Lot serial: OR); Model Catalog: 1208-4H1 (Lot serial: OR); Model Catalog: 10127-4H1 (Lot serial: OR); Model Catalog: RC41-12027 (Lot serial: ALL STARTING WITH THE LETTER Y); Model Catalog: 9054-4S2 (Lot serial: ALL STARTING WITH THE LETTER Y); Model Catalog: 9053-4S2 (Lot serial: ALL STARTING WITH THE LETTER Y); Model Catalog: 9052-4S2 (Lot serial: ALL STARTING WITH THE LETTER Y); Model Catalog: 6800-503 (Lot serial: ALL STARTING WITH THE LETTER Y); Model Catalog: 5732-4H1 (Lot serial: ALL STARTING WITH THE LETTER Y); Model Catalog: 5682-409 (Lot serial: ALL STARTING WITH THE LETTER Y); Model Catalog: 1553-4H1 (Lot serial: ALL STARTING WITH THE LETTER Y); Model Catalog: 7441-4S2 (Lot serial: ALL STARTING WITH THE LETTER Y); Model Catalog: 1209-503 (Lot serial: ALL STARTING WITH THE LETTER Y); Model Catalog: