Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Supersonic imagine has identified an error in the user's guide regarding the list of recommended cleaners/disinfectants. the disinfectants recommended for the se12-3 robe are not all high level. disinfectants. the se12-3 is used in semi-critical application so it should be cleaned and high level disinfected after use.