원인
When using 54340d allset+ gold hla abc low res kit lot 038 1249868 primer mix pm085c has been shown to produce a weak false positive reaction of correct size (420 base pair pcr amplicon) in the presence of samples containing c*04 homozygote allele(s). primer mix pm085c can be found in lane 78. as a result of the false positive reaction samples containing two c*04 alleles could produce a mistype or no-type result depending on the reaction pattern corresponding to the full abc hla typing.