Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Philips healthcare has become aware that when the c-arc base rotation is moved mannually and a motorized base rotation is initiated followed by a movement stop request the geometry of the system will reset. the stand movements and table movements will not be available for approximately 90 seconds.