Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Philips has discovered through customer feedback and internal testing the following issues: 1. a failure with the image disk may cause the entire image processing chain to be blocked due to a sw design error. this causes image data to be lost and eventually loss of live imaging functionality. 2. a filament error that may occur during the execution of an xperct. when the filament error occurs the run is aborted. after abort new x-ray runs can be started immediately without system restart.