원인
Philips has identified a problem related to the hard disk firmware in allura xper systems. the hard disk itself is working firmware has an incorrect incrementing smart counter. this might result in a very high "raw read error" count which will lead to a so called "smart trip" and subsequently will take the hard disk offline. consequently the system might intermittently freeze. this can vary from several seconds up to one minute.
Model Catalog: ALLURA XPER FD10 (Lot serial: ); Model Catalog: ALLURA XPER FD20 (Lot serial: ); Model Catalog: ALLURA XPER FD20/10 (Lot serial: ); Model Catalog: ALLURA XPER FD10/10 (Lot serial: )