Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Philips has become aware that a mechanical noise/tick might occur when moving the xy-carrier over the x-rail coupling. this problem can occur when an "old" version of coupling block with m5 bolts is installed in the system.