Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Steris has become aware that the v-pro 60 software verifies expiration date of the vaprox sterilant upon completion of the sterilization cycle rather than prior to cycle initiation. verification of the vaprox expiration date at the end of the sterilization cycle will allow for the completion of one cycle with expired vaprox sterilant.