Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Baxter received reports of 8 cases of broken/damaged wheels on the aquarius. the damage to the wheels occurred during the moving of the device. the damaged wheel may result in loss of stability of the device. in the worst case the device can fall over and potentially harm a person.