ARCHITECT C 16000 - PROCESSING MODULE 의 리콜

Health Canada (via FOI)에 따르면, 해당 리콜 는 Canada 에서 ABBOTT LABORATORIES LIMITED. DIAGNOSTIC DIVISION 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    46071
  • 사례 위험등급
    II
  • 사례 시작날짜
    2017-12-13
  • 사례 국가
  • 사례 출처
    HC
  • 비고 / 경고
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • 데이터 추가 비고
  • 원인
    The base of the architect cuvette segment may become detached under specific conditions. when a cuvette segment base is detached cuvettes may be seated lower than the designed height. this may result in inadequate dispense into specific cuvettes due to the sample probe being unable to make efficient contact with the cuvette bottom.

Device

  • 모델명 / 제조번호(시리얼번호)
    Model Catalog: 09D32-05 (Lot serial: >10 NUMBERS CONTACT MFG); Model Catalog: 02P75-01 (Lot serial: >10 NUMBERS CONTACT MFG); Model Catalog: 01G46-01 (Lot serial: >10 NUMBERS CONTACT MFG)
  • 제품 설명
    ARCHITECT C 16000 - PROCESSING MODULE;ARCHITECT C4000 SYSTEM - PROCESSING MODULE;ARCHITECT c8000 - ANALYZER;ARCHITECT C4000 SYSTEM - INTEGRATED PROCESSING MODULE
  • Manufacturer

Manufacturer

  • 제조사 주소
    MISSISSAUGA
  • 제조사 모회사 (2017)
  • 제조사 의견
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.
  • Source
    HC