  • 사례 유형
    Recall
  • 사례 ID
    77952
  • 사례 위험등급
    III
  • 사례 시작날짜
    2009-07-23
  • 사례 국가
  • 사례 출처
    HC
  • 비고 / 경고
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • 데이터 추가 비고
  • 원인
    Under specific conditions (such as spills leaks or improper maintenance) on the architect i2000 or i2000sr systems the concentrated wash buffer (containing sodium azide) can come in contact with the wash zone mechanism ground straps (cables) on either of the two wash zones. the copper in the ground strap may corrode and form an unstable chemical substance (metal azide) that may be sensitive to direct pressure and impact.

Device

  • 모델명 / 제조번호(시리얼번호)
    Model Catalog: 3M74-01 (Lot serial: ); Model Catalog: 3M74-02 (Lot serial: ); Model Catalog: 08C89-01 (Lot serial: )
  • 제품 설명
    ARCHITECT i2000 AND i2000SR SYSTEMS
  • Manufacturer

Manufacturer

  • 제조사 주소
    MISSISSAUGA
  • 제조사 모회사 (2017)
  • 제조사 의견
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.
  • Source