원인
Bard is informing customers that use of the arctic sun temperature management system in excess of 2000 system hours without conducting preventative maintenance may result in failure of certain components of the system particularly in relation to the mixing pump which runs continuously during operation. failure of the mixing pump may cause the device to cease to function as intended resulting in no availability of cooling or rewarming therapy for the patient.