Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Multiple isocenter plans will load on the 4 ditc and are available for imaging and treatment but the 4ditc software does not recognize that multiple isocenters are present. acquired couch parameters can be applied to any or all fields in the treatment plan regardless of isocenter definition. this involves product versions 7.029 8.1.1.3 8.1.38.1.4.0 8.3.0.12 8.6.1 8.6.2.0 8.8.1.7.