Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
A behavior seen with varian treatment (vt) software when either connected to a siemens linear accelerator using digital mevatron interface protocols (dmip) 5 or 6 in delivery of a conventional or imrt plan with more than one field and/or when vt is installed with the couch value propagation feature. the user may not be aware that couch rotation values were updated for all user selected fields following the use of the couch value propagation feature.