Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Baxter corporation is issuing a device correction for the artis dialysis system in order to update the current software versions to new software versions. the current version allows for the possibility of resetting and continuing patient treatment without following the instructions for addressing alarm #642 "arterial chamber: level adjustment required." this will subsequently deactivate the alarm for the remaining treatment time. the new software versions prevent the possibility of continuing the treatment without following the instructions for addressing alarm #642 as written in the operator's manual.