Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
When astral ventilators are used with the internal battery as the sole power source it is important to note that failure of the internal battery can lead to cessation of therapy without the low battery or critically low battery alarm sounding even when the battery indicates that it is charged. this could put patients at risk to not getting therapy.
Model Catalog: 27003 (Lot serial: manufacturer.); Model Catalog: 27003 (Lot serial: >10 serial numbers. Contact); Model Catalog: 27919 (Lot serial: >10 serial numbers. Contact); Model Catalog: 27919 (Lot serial: manufacturer.)