원인
Beckman coulter has identified a rare potential issue with the tip present sensor within the robot of the aliquoter module. there has been one report of a misaligned sensor in an automate system which resulted in the system processing samples without pipette tips. a small number of samples were subject to cross-contamination and were sent to the error rack prior to analysis. in this particular case the system gave no external sign of a misaligned tip present sensor.