Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
If the tabletop is approximately fully extended at the head end and there is a large amount of liquid on the table surface the liquid can make its way from the foot end of the tabletop onto the top cover of the table column where it can seep into the table through a gap. if this causes electronic components to become contaminated table movements may no longer be possible.