Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Baxter corporation has issued an urgent product recall because the slide clamp orientation may have been assembled incorrectly with the slide clamp oriented distally towards the drip chamber instead of proximally towards the pall filter. this could cause the infusion set to be placed into an infusion pump improperly and cause a backflow situation.