Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
In march 2011 the material and molding process of the coolant cup (c cups) was changed. since the change beckman coulter has received complaints regarding leaking and cracked c cups. when this occurs coolant can leak onto and damage one of the printed circuit boards (pcbs) in the analyzers.