BECKMAN COULTER AU IMMUNOCHEMISTRY SYSTEM - CREATININE (ENZYMATIC) 의 리콜

Health Canada (via FOI)에 따르면, 해당 리콜 는 Canada 에서 BECKMAN COULTER CANADA L.P. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    18909
  • 사례 위험등급
    II
  • 사례 시작날짜
    2014-12-05
  • 사례 국가
  • 사례 출처
    HC
  • 비고 / 경고
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • 데이터 추가 비고
  • 원인
    N-acetyl cysteine (nac) when administered in therapeutic concentrations in excess of 500 mg/l (for the treatment of acetaminophen overdose) has been recently determined to interfere with assays for the following analytes: enzymatic creatinine osr6x204 cholesterol osr6x16 uric acid osr6x98 lactate osr6x93 and lipase osr6x30. only these assays are affected. nac is an anti-oxidant and can interfere with assays employing trinder chemical pathways i.E. where hydrogen peroxide (h2o2) is formed and subsequenly reacted with aminoantipyrine and an indicator in the presence of a peroxidase as a catalyst. nac interference has been demonstrated to occur to varying degrees depending on the analyte and therefore not all assays employing trinder chemical pathways are affected. nac interference may lead to falsely low results.

Device

  • 모델명 / 제조번호(시리얼번호)
    Model Catalog: OSR61204 (Lot serial: ALL LOTS); Model Catalog: OSR6616 (Lot serial: ALL LOTS); Model Catalog: OSR6116 (Lot serial: ALL LOTS); Model Catalog: OSR6216 (Lot serial: ALL LOTS); Model Catalog: OSR6516 (Lot serial: ALL LOTS); Model Catalog: OSR6698 (Lot serial: ALL LOTS); Model Catalog: OSR6298 (Lot serial: ALL LOTS); Model Catalog: OSR6130 (Lot serial: ALL LOTS); Model Catalog: OSR6198 (Lot serial: ALL LOTS); Model Catalog: OSR6098 (Lot serial: ALL LOTS); Model Catalog: OSR6230 (Lot serial: ALL LOTS); Model Catalog: OSR6193 (Lot serial: ALL LOTS)
  • 제품 설명
    BECKMAN COULTER AU IMMUNOCHEMISTRY SYSTEM - ENZYMATIC CREATININE
  • Manufacturer

Manufacturer