원인
Unpaired lots of au lyophilized chemistry calibrator level 1 and 2 have been distributed. calibrator level 1 and 2 must be paired when used for calibration and cannot be used in any other combination. paired calibrator lots are identified by the last two digits of the lot number. if lot numbers are not correctly paired calibration can be run using incorrect set points which may generate erroneous results.