Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Beckman coulter has identified that there is a possibility that drug interferences from administration of metamizole (dipyrone) to patients may generate false low results with the assays which are the subject of this recall. the issue was discovered on 23rd feb 2016 internally by beckman coulter.
Model Catalog: OSR6298 (Lot serial: ALL LOTS); Model Catalog: OSR6198 (Lot serial: ALL LOTS); Model Catalog: OSR6098 (Lot serial: ALL LOTS); Model Catalog: OSR6616 (Lot serial: ALL LOTS); Model Catalog: OSR6516 (Lot serial: ALL LOTS); Model Catalog: OSR6216 (Lot serial: ALL LOTS); Model Catalog: OSR6116 (Lot serial: ALL LOTS); Model Catalog: OSR6193 (Lot serial: ALL LOTS); Model Catalog: OSR60118 (Lot serial: ALL LOTS); Model Catalog: OSR61118 (Lot serial: ALL LOTS); Model Catalog: OSR66118 (Lot serial: ALL LOTS); Model Catalog: OSR61204 (Lot serial: ALL LOTS); Model Catalog: OSR6130 (Lot serial: ALL LOTS); Model Catalog: OSR6230 (Lot serial: ALL LOTS)