Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Company received reports of incorrect results generated by analyzer as a result of cuvette overflow. current system software unable to complete real-time monitoring of reaction profiles to identify impact on results occasionally yielding incorrect results without any flags or alarms to alert operator.
Model Catalog: AU640E (Lot serial: All serial numbers); Model Catalog: AU600 (Lot serial: All serial numbers); Model Catalog: AU640 (Lot serial: All serial numbers); Model Catalog: AU480 (Lot serial: All serial numbers); Model Catalog: AU680 (Lot serial: All serial numbers); Model Catalog: AU2700 (Lot serial: All Serial Numbers); Model Catalog: AU5400 (Lot serial: All serial numbers); Model Catalog: AU400E (Lot serial: All serial numbers); Model Catalog: AU400 (Lot serial: All serial numbers)